Contact Us

Home > Products > Water Treatment & Filtration > Createc > Reverse Osmosis Controller > ROC-5200 Laboratory High Purity water operation management system

Product Name
ROC-5200 Laboratory High Purity water operation management system
Products Description
Application:
It is used for Laboratory ultra-pure water producing systems: integrated RO process controller and conductivity, resistivity monitoring.

Features:
240x180 dot matrix display screen
Integrated high pure water system operation program inside, only needs to set parameters.
Automatic measure range, digital temperature compensation
The checking terminal adopt low pressure pull-up voltage, photoelectronic seclusion input
The control terminals use no contact relay.
State checking, operation control, consumables lifespan fore notice.
Perfect disinfect process, automatic operation of disinfect process after the disinfector is put in
Calibration of electrode constant, conductivity alarm setting
DC 24V power supply meets operation in damp working place.
 
Technical Data
 
Model: ROC-5200
Reverse osmosis collection point:  
raw water no-water protection, high pressure water full protection, operation reset
 
Reverse osmosis control point:
high pressure pump, cycle pump, flush valve, RO outlet valve, disinfect valve, disinfect discharge valve, DI(pure water ) outlet valve, inner cycle valve
 
Measure collection Point:
raw water conductivity collection, conductivity value collection of single stage product ion water, conductivity value collection of two stage production water, temperature value collection
 
Measure item:
conductivity/resistivity/temperature
 
Auxiliary electrode:
Measure conductivity use 1.0cm-1 electrode
Measure resistivity with 0.05 cm-1 electrode
 
Measurement range:    
raw water conductivity 0~2000ppm with 1.0cm-1 electrode
RO conductivity 0~200ppm with 1.0cm-1 electrode
DI resistivity 0~18.2MΩ.cm(with 0.05cm-1 electrode)
 
Accuracy: ≤ 1.5%(FS)
 
Temperature compensation:  
with 25 as reference temperature,conductivity automatic digital compensation , resistivity automatic analog compensation
 
Working environment: temperature: 0-50   humidity : ≤80%RH
Electrical contactor: OC door output
Contactor Contact:
AC 220V/3A Max;AC 110V/5A Max
DC24V/3A Max(under resisitity load)
 
Data renew rate: ≤1second/time
Conductance working pressure: 0~0.5Mpa
Power supply: DC 24V±4V
Outline dimension: 160mm×100mm×40mm(H×W×D)
Hole size: 57mm×106mm (H×W)
Fix mode: Inside the machine
 
 
 
??????? ???????? ?????? ?????
????? ??????
 03-6764586
?? ????
 03-6771280






Pumps

Flow and Level

Valves & Fittings

Air and Gas Handling & Managment

Water Treatment & Filtration

- Siemens

- ATG UV Technology

- CLEAR LAKE Filtration

- Createc

- Pentair

- GE

- EUROTROL

Orbital Welding

Stock Products For sale

 

 

 

 

 

© כל הזכויות שמורות לחברת סמל בע"מ

 

Home   |    Products   |    About us   |    News   |    Companies & Downloads   |    Sitemap   |    Terms & Conditions   |    Contact Us   |    Privacy Policy